Philips North America Llc Recall 89375

Description: Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031

Philips North America Llc Recall 89375 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0725-2022
Event ID89375
Event DescriptionPhilips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031
Product TypeDevices
DistributionUS Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sri Lanka Switzerland Taiwan Thailand Turkey United Kingdom Virgin Islands (U.S.)
Quantity1 unit
Recall ReasonDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Device Classification20220316
Device Code InfoS/N: OUS 109
Center Classification Date20220304
Recall Initiation Date20211124
Recalling FirmPhilips North America Llc