Ortho-Clinical Diagnostics, Inc. Recall 89362

Description: VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.

Ortho-Clinical Diagnostics, Inc. Recall 89362 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0735-2022
Event ID89362
Event DescriptionVITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
Product TypeDevices
DistributionDomestic distribution nationwide. Foreign distribution to Canada.
Quantity4125 units
Recall ReasonThe lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.
Device Classification20220316
Device Code InfoUDI 10758750006786 Lot 1039734
Center Classification Date20220308
Recall Initiation Date20211210
Recalling FirmOrtho-Clinical Diagnostics, Inc.
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit