Product Safety Recalls

Product Recall Tracker

NeuMoDx Molecular Inc Recall 89675

Description: NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 assay strips in individual pouches. Box, pouch and strips all labeled with product name, ref #, lot #, and expiration date.

NeuMoDx Molecular Inc Recall 89675 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0738-2022
Event ID89675
Event DescriptionNeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 assay strips in individual pouches. Box, pouch and strips all labeled with product name, ref #, lot #, and expiration date.
Product TypeDevices
DistributionUS Nationwide distribution in the states of FL, MD, MI, NY.
Quantity102 boxes
Recall ReasonA risk for product contamination was identified for the affected product lot. If used, there is an increased potential for SARS-CoV-2 false positive results when using the product.
Device Classification20220316
Device Code InfoLot: 116305 S/L-02 (Sublot 02),UDI: 10814278020830
Center Classification Date20220308
Recall Initiation Date20220121
Recalling FirmNeuMoDx Molecular Inc