Product Safety Recalls

Product Recall Tracker

Covidien Llc Recall 86886

Description: Covidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3L

Covidien Llc Recall 86886 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0748-2021
Event ID86886
Event DescriptionCovidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3L
Product TypeDevices
DistributionNationwide Foreign: Canada, Albania Armenia Australia Austria Belarus Belgium Bosnia And Herzegovina Cyprus Czech Republic Egypt Estonia France Germany Greece Hungary Iceland Israel Italy Japan Jordan Korea Kuwait Latvia Libyan Arab Jamahiriya Lithuania Malaysia Martinique Morocco Myanmar, Netherlands Norway Pakistan Poland Russian Federation Serbia Singapore Slovakia Spain Sweden Switzerland Taiwan Turkey United Arab Emirates United Kingdom Viet Nam
Quantity5316 units US; 1209 units OUS
Recall ReasonMay contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis
Device Classification20210106
Device Code InfoLot Numbers: P0B1335MY, P0D0170MY, P0D1445M, 0F0377MY, P0B1336MY,P0D1444MY, P0E1211MY, P0F0606MY. UDI: 10884521043152, 20884521043159
Center Classification Date20201229
Recall Initiation Date20201123
Recalling FirmCovidien Llc