Olympus Corporation of the Americas Recall 93649
Description: SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS
Olympus Corporation of the Americas Recall 93649 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0748-2024 |
Event ID | 93649 |
Event Description | SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America. |
Quantity | 82,200 total units (5 per box) |
Recall Reason | Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient. |
Device Classification | 20240124 |
Device Code Info | Models UDI/GTIN TFL-FBX150S 00821925043916 TFL-FBX200S 00821925043930 TFL-FBX365S 00821925043992 TFL-FBX550S 00821925044036 TFL-FBX940S 00821925044074 TFL-FBX150BS 00821925043879 TFL-FBX200BS 00821925043978 All lots affected |
Center Classification Date | 20240118 |
Recall Initiation Date | 20231201 |
Recalling Firm | Olympus Corporation of the Americas |
Initial Notification | Letter |
Similar To |