Cardinal Health 200, LLC Recall 93628
Description: CardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510
Cardinal Health 200, LLC Recall 93628 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0754-2024 |
Event ID | 93628 |
Event Description | CardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510 |
Product Type | Devices |
Distribution | US, Canada, Australia and New Zealand |
Quantity | 3505730 units |
Recall Reason | The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister. |
Device Classification | 20240124 |
Device Code Info | UDI/DI 50630140152649 (case), Lot Numbers: J308-602, J308-605, J308-609, J308-623, J308-624, J308-625, J308-628, J308-629, J308-630, J308-632, J308-633, J308-634, J308-635, J308-636, J308-637, J308-640, J308-642, J308-607, J308-643, J309-603, J309-607, J309-623, J309-627, J309-630, J309-632, J309-635, J309-636, J309-637, J310-601, J310-609, J310-621, J310-622, J310-626, J310-627, J310-641, J310-642, J310-643, J310-644 |
Center Classification Date | 20240118 |
Recall Initiation Date | 20231221 |
Recalling Firm | Cardinal Health 200, LLC |
Initial Notification | Letter |