Product Safety Recalls

Product Recall Tracker

American Contract Systems, Inc. Recall 93656

Description: Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LavH, kit number BBLV21C; b) LAP HYST 247540, kit number BFHY75U

American Contract Systems, Inc. Recall 93656 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0831-2024
Event ID93656
Event DescriptionCustom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LavH, kit number BBLV21C; b) LAP HYST 247540, kit number BFHY75U
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity144 units
Recall ReasonDuring an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Device Classification20240207
Device Code Infoa) BBLV21C, UDI/DI none, Lot Numbers: 686221, exp. 11/10/2023; 791221, exp. 7/28/2023; 841221, exp. 6/8/2023; b) BFHY75U, UDI/DI 00191072169839, Lot Numbers: 878231, exp. 5/2/2024; 933231, exp. 3/8/2024; 987231, exp. 11/30/2023; 659221, exp. 12/7/2023; 689221, exp. 11/7/2023
Center Classification Date20240126
Recall Initiation Date20231129
Recalling FirmAmerican Contract Systems, Inc.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.