Fresenius Medical Care Holdings, Inc. Recall 86855
Description: NaturaLyte Liquid Acid concentrate drums, Model Number 13-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Fresenius Medical Care Holdings, Inc. Recall 86855 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0823-2021 |
Event ID | 86855 |
Event Description | NaturaLyte Liquid Acid concentrate drums, Model Number 13-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine. |
Product Type | Devices |
Distribution | US Nationwide distribution. |
Quantity | 51 drums |
Recall Reason | The sodium and/or calcium levels in these concentrates were below the manufacturing specifications. |
Device Classification | 20210120 |
Device Code Info | Lot Numbers: 20LXAC034 20LXAC069 20NXAC040 |
Center Classification Date | 20210111 |
Recall Initiation Date | 20201107 |
Recalling Firm | Fresenius Medical Care Holdings, Inc. |
Initial Notification | Letter |
Similar To |