Product Safety Recalls

Product Recall Tracker

Baxter Healthcare Corporation Recall 89598

Description: Liko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model numbers 3156087 and P3156087.

Baxter Healthcare Corporation Recall 89598 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0829-2022
Event ID89598
Event DescriptionLiko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model numbers 3156087 and P3156087.
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, ND, NE, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. There was government distribution but no military distribution. The countries of Austria, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, Ireland, Japan, Jordan, Latvia, Netherlands, Portugal, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Kingdom, and Uruguay.
Quantity1,399 twinbars
Recall ReasonAn incorrect center bolt was used in the assembly of the product which could result in a patient fall during use.
Device Classification20220406
Device Code InfoManufactured before 28OCT2021, UDI 00887761982468
Center Classification Date20220328
Recall Initiation Date20220221
Recalling FirmBaxter Healthcare Corporation
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.