Siemens Healthcare Diagnostics, Inc. Recall 89697

Description: Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension Vista System. The Dimension Vista Magnesium Flex Reagent (MG) method is an In vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma, and urine on the Dimension Vista System. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypomagnesemia (abnormally high plasma levels of magnesium).

Siemens Healthcare Diagnostics, Inc. Recall 89697 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0830-2022
Event ID89697
Event DescriptionSiemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension Vista System. The Dimension Vista Magnesium Flex Reagent (MG) method is an In vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma, and urine on the Dimension Vista System. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypomagnesemia (abnormally high plasma levels of magnesium).
Product TypeDevices
DistributionWorldwide distribution: US (nationwide) 47 States including: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY) and OUS countries of: Australia, Canada and Netherlands.
Quantity14220 (US 13305, OUS 915)
Recall ReasonDimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may lead to repeat and follow-up testing.
Device Classification20220406
Device Code Info[lot 21253BB, UDI: 0084276801569421253BB22091010445158840], [lot 21263BA UDI: 0084276801569421263BA22092010445158840], [lot 21335AA, UDI: 0084276801569421335AA22120110445158840]
Center Classification Date20220328
Recall Initiation Date20220128
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Initial Notification Letter