Product Safety Recalls

Product Recall Tracker

XTANT Medical Holdings, Inc Recall 93738

Description: Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

XTANT Medical Holdings, Inc Recall 93738 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0846-2024
Event ID93738
Event DescriptionCortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.
Product TypeDevices
DistributionUS distribution in MI.
Quantity6
Recall ReasonSpinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
Device Classification20240207
Device Code InfoUDI-DI: 00810076564005, Lot: AP23000143
Center Classification Date20240129
Recall Initiation Date20231128
Recalling FirmXTANT Medical Holdings, Inc
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.