The Magstim Company Limited Recall 87063

Description: Neurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.

The Magstim Company Limited Recall 87063 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0869-2021
Event ID87063
Event DescriptionNeurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.
Product TypeDevices
DistributionMN, NC, and NY
Quantity7 units
Recall ReasonA small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.
Device Classification20210127
Device Code InfoSerial Numbers: 2507, 2511, 2514, 2515, 2522, 2527, 2533, 2534, 2537
Center Classification Date20210115
Recall Initiation Date20201208
Recalling FirmThe Magstim Company Limited