Medtronic Neurosurgery Recall 87069
Description: REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.
Medtronic Neurosurgery Recall 87069 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0872-2021 |
Event ID | 87069 |
Event Description | REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity. |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IL, KS, KY, LA, MA, MI, MO, NV, NM, NY, ND, OH, OK, TN, VA and the countries of Albania, Australia, Brazil, Cyprus, Ireland, Mexico, Qatar, Romania, Russian Federation, Serbia, South Africa, Spain, United Kingdom. |
Recall Reason | Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction |
Device Classification | 20210127 |
Device Code Info | Lot # 0219281777 |
Center Classification Date | 20210115 |
Recall Initiation Date | 20201216 |
Recalling Firm | Medtronic Neurosurgery |
Initial Notification | Letter |
Similar To |