Volcano Corp Recall 89698
Description: MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems
Volcano Corp Recall 89698 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0917-2022 |
Event ID | 89698 |
Event Description | MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems |
Product Type | Devices |
Distribution | US: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO, KY, LA, DE, MA, MT, MO, VA, AL, NY, WV, SC, IN, GA, IA, OH, UT, NH, MS, TN, OK, ID, MD, HI, RI, WY, DC, WA, KS, NE, ND, CT, VT. OUS: SK, IT, DK, GB, DE, SE, IN, KW, QA, PL, NL, CL, JP, BE, AU, FR, PF, PT, NZ, CA, ES, PK, CY, AT, IL, IE, CH, GR, SA, PS, LB, HK, IS, TH, AR, OM, AE, NO, ID, RO, RS, MM, ZA, SG |
Quantity | 1358 |
Recall Reason | Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets. |
Device Classification | 20220420 |
Device Code Info | MM-TSMs with bottom protective beams |
Center Classification Date | 20220412 |
Recall Initiation Date | 20220203 |
Recalling Firm | Volcano Corp |
Initial Notification | Letter |
Similar To |