OMNIlife science Inc. Recall 87062
Description: SERF Novae Dual Mobility Cup Insert 49/28-ApeX-LNK Poly Liner. Intended for use with the Apex Modular Acetabular Cup for total hip replacement procedures. Product Code: H6-24928
OMNIlife science Inc. Recall 87062 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0922-2021 |
Event ID | 87062 |
Event Description | SERF Novae Dual Mobility Cup Insert 49/28-ApeX-LNK Poly Liner. Intended for use with the Apex Modular Acetabular Cup for total hip replacement procedures. Product Code: H6-24928 |
Product Type | Devices |
Distribution | AL, ID, MD, NY, OK |
Quantity | 2 units |
Recall Reason | Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery |
Device Classification | 20210203 |
Device Code Info | Lot Number: 28572 |
Center Classification Date | 20210126 |
Recall Initiation Date | 20201221 |
Recalling Firm | OMNIlife science Inc. |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |