Product Safety Recalls

Product Recall Tracker

AtriCure, Inc. Recall 93672

Description: Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053

AtriCure, Inc. Recall 93672 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0927-2024
Event ID93672
Event DescriptionCryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053
Product TypeDevices
DistributionUS Nationwide. Brazil, Canada, Japan, Taiwan
Quantity35
Recall ReasonAtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
Device Classification20240207
Device Code InfoUDI-DI: 10840143903038; Lot Numbers: 129673, 129674, 129882, 130314, 130750, 131614, 132669, 133672, 133673, 134270
Center Classification Date20240131
Recall Initiation Date20231128
Recalling FirmAtriCure, Inc.