Brainreader ApS Recall 86995
Description: Neuroreader Medical Image Processing Software - Product Usage: intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on MR images.
Brainreader ApS Recall 86995 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0934-2021 |
Event ID | 86995 |
Event Description | Neuroreader Medical Image Processing Software - Product Usage: intended to automate the current manual process of identifying, labeling, and quantifying the volume of segmentable brain structures identified on MR images. |
Product Type | Devices |
Distribution | US Nationwide distribution including in the states of UT, PA, FL, AL, CA, MN, MD, MO, VA, DC, IL, TX, TN, MI, NY, IN, WA, WI, GA, MA. |
Quantity | 43 sites have had access |
Recall Reason | The Neuroreader has been distributed with a reference normative database that is different from the version reviewed and cleared by FDA. |
Device Classification | 20210210 |
Device Code Info | Version 2.4 |
Center Classification Date | 20210129 |
Recall Initiation Date | 20201214 |
Recalling Firm | Brainreader ApS |
Initial Notification | |
Similar To |