Graphic Controls Acquisition Corporation Recall 89810
Description: CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US
Graphic Controls Acquisition Corporation Recall 89810 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0936-2022 |
Event ID | 89810 |
Event Description | CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US |
Product Type | Devices |
Distribution | US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT. |
Quantity | 560 units |
Recall Reason | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation. |
Device Classification | 20220420 |
Device Code Info | All lots UDI: 03700506307340 |
Center Classification Date | 20220414 |
Recall Initiation Date | 20220315 |
Recalling Firm | Graphic Controls Acquisition Corporation |