The Binding Site Group, Ltd. Recall 86926
Description: Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A
The Binding Site Group, Ltd. Recall 86926 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0943-2021 |
Event ID | 86926 |
Event Description | Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A |
Product Type | Devices |
Distribution | U.S.: CO, FL, IA, MA, NC, NJ, NY, OH, SD, and UT. O.U.S.: Not provided |
Quantity | 213 kits in the U.S. |
Recall Reason | Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values. |
Device Classification | 20210210 |
Device Code Info | Model Number: LK016.M.OPT.A UDI: 05051700019866 Lot Number: 454763 |
Center Classification Date | 20210201 |
Recall Initiation Date | 20201118 |
Recalling Firm | The Binding Site Group, Ltd. |
Initial Notification | |