Product Safety Recalls

Product Recall Tracker

Mar-Med Co Recall 87009

Description: Tourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.

Mar-Med Co Recall 87009 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0945-2021
Event ID87009
Event DescriptionTourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Germany, Greece, Italy, Japan, Netherlands, Singapore, Spain, Sweden, Switzerland, and United Kingdom.
Quantity52,000
Recall ReasonA failure can occur if the user attempts to remove the device by partially snipping the edge and pulling on it. The device may tear into two pieces with a portion remaining in place, which can continue to constrict the vessels. This type of failure can result in digit ischemia.
Device Classification20210210
Device Code InfoAll lots
Center Classification Date20210202
Recall Initiation Date20201228
Recalling FirmMar-Med Co
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.