Becton Dickinson & Co. Recall 87178
Description: BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.
Becton Dickinson & Co. Recall 87178 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0954-2021 |
Event ID | 87178 |
Event Description | BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider. |
Product Type | Devices |
Distribution | US nationwide distribution. |
Quantity | 938 kits |
Recall Reason | Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may be mislabeled. This could result in a kit containing two pouches of N1 master mix or two pouches of N2 master mix, instead of one pouch of each. If laboratory personnel do not notice that both pouches contain tubes with the same color foil seal and use two of the same tubes, there is a low potential for a false negative SARS-CoV-2 result. A false negative SARS-CoV-2 result could lead to an increased risk of disease progression for the patient and additional spread of SARS-CoV-2 if the patient is not quarantined. |
Device Classification | 20210217 |
Device Code Info | Catalog Number: 444213, Lot Numbers: K20-507, K20-508, K20-509, K20-516, K20-517, K21-020, K21-021. GTIN Number: 00382904442132 |
Center Classification Date | 20210205 |
Recall Initiation Date | 20210107 |
Recalling Firm | Becton Dickinson & Co. |
Initial Notification | Letter |