Siemens Medical Solutions USA, Inc Recall 89860

Description: Artis zee floor MN, Model Number 10094142

Siemens Medical Solutions USA, Inc Recall 89860 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0971-2022
Event ID89860
Event DescriptionArtis zee floor MN, Model Number 10094142
Product TypeDevices
DistributionUS Nationwide.
Quantity24 systems
Recall ReasonIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Device Classification20220427
Device Code InfoUDI: 4056869010090; Serial Numbers: 140421 140115 140101 140205 140213 140204 140209 140219 140224 140218 140226 140221 140220 140206 140105 140423 140406 140422 140102 140214 140109 140202 140424 140114
Center Classification Date20220421
Recall Initiation Date20220223
Recalling FirmSiemens Medical Solutions USA, Inc
Initial Notification Letter