Siemens Medical Solutions USA, Inc Recall 89870
Description: Artis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327600
Siemens Medical Solutions USA, Inc Recall 89870 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0985-2022 |
Event ID | 89870 |
Event Description | Artis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327600 |
Product Type | Devices |
Distribution | US Nationwide distribution. |
Quantity | 94 U.S. |
Recall Reason | Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movement s function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed |
Device Classification | 20220427 |
Device Code Info | S/N: 180062 180054 180019 180065 180056 180057 180041 180421 180317 180034 180072 180349 180367 180382 180060 180337 180070 180048 180083 180369 180370 180371 180372 180345 180315 180343 180080 180301 180305 180067 180330 180040 180379 180386 180360 180333 180029 180352 180336 180407 180412 180400 180397 180316 180068 180363 180084 180418 180334 180018 180350 180376 180354 180046 180047 180335 180085 180035 180339 180357 180358 180361 180362 180081 180423 180331 180411 180414 180381 180021 180342 180321 180309 180356 180378 180393 180368 180413 180341 180323 180329 180043 180049 180377 180077 180086 180053 180055 180015 180014 180066 180082 180037 180373 UDI: 04056869063317 |
Center Classification Date | 20220421 |
Recall Initiation Date | 20220225 |
Recalling Firm | Siemens Medical Solutions USA, Inc |
Initial Notification | Letter |
Similar To |