Product Safety Recalls

Product Recall Tracker

Getinge Usa Sales Inc Recall 93736

Description: Flow-c Anesthesia System, model 6887700

Getinge Usa Sales Inc Recall 93736 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0990-2024
Event ID93736
Event DescriptionFlow-c Anesthesia System, model 6887700
Product TypeDevices
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
Quantity14 (US)
Recall ReasonPotential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Device Classification20240214
Device Code InfoUDI-DI: 07325710009765, serial numbers (US): 5024, 5025, 5026, 5027, 5028, 5043, 5044, 5045, 5046, 5047, 5048, 5049, 5050, 5078.
Center Classification Date20240206
Recall Initiation Date20231211
Recalling FirmGetinge Usa Sales Inc
Initial Notification Letter