Product Safety Recalls

Product Recall Tracker

Olympus Corporation of the Americas Recall 93792

Description: Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.

Olympus Corporation of the Americas Recall 93792 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1009-2024
Event ID93792
Event DescriptionWater Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.
Product TypeDevices
DistributionDomestic US distribution nationwide.
Quantity877 MAJ-2318, 119 OER-Elite, 3 OER-MINI
Recall ReasonWater filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
Device Classification20240214
Device Code InfoUDI-DI 04953170401473 Lots 36A, 37A, 38A Affected product may have also been packaged within OER boxes (OER-Elite and OER-MINI)
Center Classification Date20240207
Recall Initiation Date20240116
Recalling FirmOlympus Corporation of the Americas
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.