Instrumentation Laboratory Recall 89894
Description: ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097
Instrumentation Laboratory Recall 89894 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1014-2022 |
Event ID | 89894 |
Event Description | ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, El Salvador, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, S. Tome, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad, Tobago, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam. |
Quantity | 2725 units US: 5,572 units OUS.UPDATE: 2727 units US; 5277 units OUS |
Recall Reason | Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe |
Device Classification | 20220504 |
Device Code Info | All Serial Numbers. Trade Name Part Number UDI ACL TOP 300 CTS 00000280060 08426950556916; "ACL TOP 300 CTS (Refurbished)" 00000280060R 08426950928027; ACL TOP 350 CTS 00000280065 08426950784081; ACL TOP 500 CTS 00000280040 08426950453499; "ACL TOP 500 CTS (Refurbished)" 00000280040R 08426950928003; ACL TOP 550 CTS 00000280045 08426950729242; ACL TOP 970 CL 00000280097 08430793045476. |
Center Classification Date | 20220428 |
Recall Initiation Date | 20220222 |
Recalling Firm | Instrumentation Laboratory |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |