Product Safety Recalls

Product Recall Tracker

Becton Dickinson & Co. Recall 89896

Description: BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Becton Dickinson & Co. Recall 89896 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1018-2022
Event ID89896
Event DescriptionBD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10
Product TypeDevices
DistributionWorldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.
Quantity15 systems
Recall ReasonWhen updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.
Device Classification20220511
Device Code InfoAll systems running BD Synapsys version 4.10; UDI: 00382904441500; Serial Numbers: SL00530 SL00515 SL00527 SL00502 SL00540 SL00564 SL00549 SL00509 SL00003 SL00004 SL00010 SL00001 SL00005 SL00002 SL000015 SL01010 SL01022
Center Classification Date20220429
Recall Initiation Date20220309
Recalling FirmBecton Dickinson & Co.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.