Abiomed, Inc. Recall 93746

Description: Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 100080 and 0048-0045 - Impella CP with SmartAssist -- ***Added 2/20/24*** (10) 4413 and 004413 - Impella 2.5 Set (11) 5040 and 005040 - Impella 2.5 IMC Pump Set EU (12) 5060 and 005060 - Impella 5.0 IMC Pump Set EU (13) 5064 and 005064 - Impella 5.0 IMC Pump Set Canada (14) 1000115 - Impella CP Pump set, APAC (15) 1000211 - Impella 5.5 Smart Assist Set, JP (16) 1000302 - Impella CP with SmartAssist APAC (17) 1000361 - Impella 5.5 Set AU (18) 1000402 - Impella CP Smart Assist Set (19) 0046-0011 - Impella RP Pump Set, EU (20) 0046-0026 - Impella 5.0 Pump Set ROW (21) 0046-0037 - Impella 5.0 Pump Set APAC (22) 0046-0039 - Impella RP Set APAC (23) 004690-AU - Impella 5.0 Pump Set AU (24) 0048-0002 - Impella CP Pump Set, EU (25) 0048-0002-BR - Impella CP Pump Set BR (26) 0048-0004 - Impella CP Pump Set, Canada (27) 0048-0014 - Impella Smart Assist Set, EU (28) 0048-0024-JP - Impella CP Smart Assist Set, JP (29) 0048-0044 - Impella CP Smart Assist Set, Canada (30) 0048-0047- Impella CP Smart Assist Set APAC (31) 005048-JP - Impella 2.5 Pump Set, Japan (32) 005066-JP - Impella 5.0 Pump Set, Japan (33) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (34) 0550-0004 - Impella 5.5 with SmartAssist Set, CA

Abiomed, Inc. Recall 93746 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1074-2024
Event ID93746
Event DescriptionImpella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 100080 and 0048-0045 - Impella CP with SmartAssist -- ***Added 2/20/24*** (10) 4413 and 004413 - Impella 2.5 Set (11) 5040 and 005040 - Impella 2.5 IMC Pump Set EU (12) 5060 and 005060 - Impella 5.0 IMC Pump Set EU (13) 5064 and 005064 - Impella 5.0 IMC Pump Set Canada (14) 1000115 - Impella CP Pump set, APAC (15) 1000211 - Impella 5.5 Smart Assist Set, JP (16) 1000302 - Impella CP with SmartAssist APAC (17) 1000361 - Impella 5.5 Set AU (18) 1000402 - Impella CP Smart Assist Set (19) 0046-0011 - Impella RP Pump Set, EU (20) 0046-0026 - Impella 5.0 Pump Set ROW (21) 0046-0037 - Impella 5.0 Pump Set APAC (22) 0046-0039 - Impella RP Set APAC (23) 004690-AU - Impella 5.0 Pump Set AU (24) 0048-0002 - Impella CP Pump Set, EU (25) 0048-0002-BR - Impella CP Pump Set BR (26) 0048-0004 - Impella CP Pump Set, Canada (27) 0048-0014 - Impella Smart Assist Set, EU (28) 0048-0024-JP - Impella CP Smart Assist Set, JP (29) 0048-0044 - Impella CP Smart Assist Set, Canada (30) 0048-0047- Impella CP Smart Assist Set APAC (31) 005048-JP - Impella 2.5 Pump Set, Japan (32) 005066-JP - Impella 5.0 Pump Set, Japan (33) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (34) 0550-0004 - Impella 5.5 with SmartAssist Set, CA
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.
Quantity92515
Recall ReasonNew warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.
Device Classification20240221
Device Code InfoUDI-DI (1) 00813502011081 (2) 00813502011180 (3) 00813502011227 (4) 00813502011029 (5) 00813502011869 (6) 00813502012811 (7) 00813502011531 and 00813502012828 (8) 00813502011388 (9) 00813502012279 and 00813502011876 All products are affected by IFU Update ***Added 2/20/24*** (10) 813502010947 (11) 4260113630136 (12) 4260113630174 (13) 813502011197 (14) 813502010473 (15) 813502012453 (16) 813502012767 (17) 813502012927 (18) 813502012873 (19) 4260113630273 (20) 813502011821 (21) 813502011937 (22) 813502011951 (23) N/A (24) 4260113630242 (25) 813502011777 (26) 813502011265 (27) 4260113630280 (28) 813502011609 (29) 813502012200 (30) 813502011944 (31) 813502010046 (32) 813502010053 (33) 813502011630 (34) 813502010466
Center Classification Date20240212
Recall Initiation Date20231227
Recalling FirmAbiomed, Inc.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.