Product Safety Recalls

Product Recall Tracker

SEDECAL SA Recall 93651

Description: MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007

SEDECAL SA Recall 93651 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1083-2024
Event ID93651
Event DescriptionMobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007
Product TypeDevices
DistributionUS Nationwide distribution. There was government distribution but no military distribution.
Quantity851 devices
Recall ReasonWhile cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
Device Classification20240221
Device Code InfoModel 712001 - UDI-DI 00884838028586, all serial numbers; Model 712002 - UDI-DI 00884838082373, all serial numbers; Model 712004 - UDI-DI 08436046002173, all serial numbers; Model 712006 - UDI-DI 00884838091504, serial numbers: SN20410556 and SN21410184; Model 712007 - UDI-DI)08436046002357, serial numbers: 26, 27, 19410038, 19410044, 19410046, 21410261, 21410266, 21410280, 21410288, 21410290, 21410291, 21410293, 21410294, 21410339, 21410340, 21410341, 21410342, 21410345, 22410022, 22410023, SN0410157, SN1410049, SN19410034, SN19410035, SN19410036, SN19410037, SN19410071, SN19410092, SN19410093, SN19410094, SN19410097, SN19410098, SN19410099, SN19410100, SN19410101, SN19410102, SN19410115, SN19410116, SN19410123, SN19410127, SN19410129, SN19410135, SN19410136, SN19410137, SN20410029, SN20410030, SN20410032, SN20410033, SN20410034, SN20410044, SN20410045, SN20410049, SN20410060, SN20410063, SN20410064, SN20410065, SN20410066, SN20410071, SN20410072, SN20410073, SN20410074, SN20410081, SN20410082, SN20410083, SN20410084, SN20410093, SN20410094, SN20410095, SN20410096, SN20410097, SN20410115, SN20410116, SN20410117, SN20410118, SN20410119, SN20410120, SN20410154, SN20410171, SN20410172, SN20410202, SN20410218, SN20410219, SN20410242, SN20410260, SN20410261, SN20410279, SN20410280, SN20410288, SN20410289, SN20410290, SN20410291, SN20410298, SN20410299, SN20410300, SN20410313, SN20410321, SN20410334, SN20410360, SN20410367, SN20410373, SN20410395, SN20410396, SN20410428, SN20410472, SN20410473, SN20410482, SN20410555, SN20410556, SN20410557, SN20410566, SN20410567, SN20410570, SN21410017, SN21410034, SN21410035, SN21410048, SN21410060, SN21410061, SN21410062, SN21410081, SN21410093, SN21410094, SN21410094, SN21410099, SN21410103, SN21410104, SN21410105, SN21410117, SN21410118, SN21410119, SN21410130, SN21410131, SN21410132, SN21410133, SN21410149, SN21410150, SN21410153, SN21410154, SN21410155, SN21410169, SN21410170, SN21410171, SN21410172, SN21410187, SN21410188, SN21410189, SN21410190, SN21410200, SN21410201, SN21410203, SN21410204, SN21410205, SN21410206, SN21410207, SN21410208, SN21410212, SN21410213, SN21410214, SN21410232, SNZY20410009, SNZY20410011, SNZY20410014, and SNZY204110015.
Center Classification Date20240214
Recall Initiation Date20211014
Recalling FirmSEDECAL SA
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.