Beckman Coulter Inc. Recall 90087

Description: MicroScan WalkAway-40 plus Instrument REF B1018-283

Beckman Coulter Inc. Recall 90087 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1083-2022
Event ID90087
Event DescriptionMicroScan WalkAway-40 plus Instrument REF B1018-283
Product TypeDevices
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The countries of Belgium, Brazil, Colombia, Germany, Hong Kong, Indonesia, Italy, Japan, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, United Arab Emirates, Vietnam and Yemen.
Quantity22 systems
Recall ReasonWalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Device Classification20220525
Device Code InfoCatalog Number: B1018-283 UDI-DI Code: 1509959065866 Serial Numbers: 34010003 34010004 34010005 34010006 34010007 34010008 34010009 34010010 34010011 34010012 34010013 34010014 34010015 34010016 34010017 34010018 34010019 34010020 34010021 34010022 34010023 34010024
Center Classification Date20220513
Recall Initiation Date20191021