Bio-Rad Laboratories, Inc. Recall 90026
Description: BioPlex 2200 REF 12000670 ToRC IgM Reagent Pack
Bio-Rad Laboratories, Inc. Recall 90026 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1096-2022 |
Event ID | 90026 |
Event Description | BioPlex 2200 REF 12000670 ToRC IgM Reagent Pack |
Product Type | Devices |
Distribution | Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ, CA, FL, HI, IA, IL, MI, MN, NE, NY, TN, TX, and WA. The countries of Czech Republic, France, Germany, and Italy. |
Quantity | 348 units |
Recall Reason | Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack. |
Device Classification | 20220525 |
Device Code Info | Model/Catalog Number: 12000670 UDI-DI (GTIN) Code: 03610520992715 Lot Number: 301388 |
Center Classification Date | 20220516 |
Recall Initiation Date | 20220414 |
Recalling Firm | Bio-Rad Laboratories, Inc. |
Initial Notification | Letter |
Similar To |