Product Safety Recalls

Product Recall Tracker

Wishbone Medical, Inc. Recall 91526

Description: WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS 10-702-095115-0 Short 10-702-115150-0 Medium 10-702-140200 Long 10-702-190300-0

Wishbone Medical, Inc. Recall 91526 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1102-2023
Event ID91526
Event DescriptionWishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS 10-702-095115-0 Short 10-702-115150-0 Medium 10-702-140200 Long 10-702-190300-0
Product TypeDevices
DistributionUS Nationwide distribution in the state of South Carolina.
Quantity195
Recall ReasonFirm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
Device Classification20230215
Device Code InfoXXS, Part No. 10-702-080095-0 UDI-DI B5621070208009500 Lot 1048176, 1048221 XS, Part No. 10-702-095115-0 UDI-DI B5621070209511500 Lot 1048177 Short, Part No. 10-702-115150-0 UDI-DI B5621070211515000 Lot 1048178, 1048222, 1048223 Medium, Part No. 10-702-140200 UDI-DI B562107021402000 Lot 1048179 Long, Part No. 10-702-190300-0 UDI-DI B5621070219030000 Lot 1048180
Center Classification Date20230207
Recall Initiation Date20230104
Recalling FirmWishbone Medical, Inc.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.