Draegar Medical Systems, Inc. Recall 87176
Description: Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.
Draegar Medical Systems, Inc. Recall 87176 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1104-2021 |
Event ID | 87176 |
Event Description | Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals. |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lithuania, Malaysia, Mexico, Monaco, Namibia, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, United Kingdom, Utd.Arab.Emir., Vietnam. |
Quantity | 19767 systems (Domestic: 3377, Foreign 16,390) |
Recall Reason | The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring. |
Device Classification | 20210217 |
Device Code Info | Model Number - MS25510, MS25520, MS26372 Revisions 24 to 29 |
Center Classification Date | 20210210 |
Recall Initiation Date | 20201223 |
Recalling Firm | Draegar Medical Systems, Inc. |
Initial Notification | Letter |