BEMER INT. AG Recall 87094
Description: Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)
BEMER INT. AG Recall 87094 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1105-2021 |
Event ID | 87094 |
Event Description | Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set) |
Product Type | Devices |
Distribution | US |
Quantity | 197,724 units |
Recall Reason | There potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker. |
Device Classification | 20210217 |
Device Code Info | All Lot Numbers |
Center Classification Date | 20210210 |
Recall Initiation Date | 20210113 |
Recalling Firm | BEMER INT. AG |
Initial Notification | Letter |
Similar To |