Arrow International Inc Recall 87092

Description: Arrow AGB+ Multi-Lumen CVC Kit 7Fr 3 Lumen- Intended for venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. Product Code: CDC-45703-1A

Arrow International Inc Recall 87092 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1106-2021
Event ID87092
Event DescriptionArrow AGB+ Multi-Lumen CVC Kit 7Fr 3 Lumen- Intended for venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. Product Code: CDC-45703-1A
Product TypeDevices
DistributionUS Nationwide
Quantity1,705 eaches
Recall ReasonThe lidstock / information card on the kits correctly states the product code as CDC-45703-1A. The banner card on the side of the kit may incorrectly state the product code as CDC-15703-1A and not identify the product as a coated ARROWg+ard Blue PLUS (AGB+) antimicrobial catheter which may cause a minor delay during use and placement of a coated catheter in patients with known sensitivity to chlorohexidine, silver sulfadiazine or sulfa could result in an allergic reaction
Device Classification20210217
Device Code InfoLot Code: 13F20E0173
Center Classification Date20210210
Recall Initiation Date20210106
Recalling FirmArrow International Inc