Medicrea International Recall 95957
Description: UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries.
Medicrea International Recall 95957 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1118-2025 |
Event ID | 95957 |
Event Description | UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries. |
Product Type | Devices |
Distribution | U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Argentina Australia Belgium Brazil Bulgaria Chile Czech Republic Denmark Finland France Germany Greece Iceland Ireland Israel Italy Luxembourg Malaysia Mexico Monaco Morocco Netherlands New Zealand Poland Portugal Singapore South Korea Spain Sweden Switzerland United Kingdom |
Quantity | 1,548 users |
Recall Reason | Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification. |
Device Classification | 20250212 |
Device Code Info | UNiD HUB is not a medical device and thus does not have a CFN or GTIN code. UNiD Spine Analyzer Product Number SW3002 UDI-DI code: 03613720286929 No lot numbers |
Center Classification Date | 20250206 |
Recall Initiation Date | 20241203 |
Recalling Firm | Medicrea International |
Initial Notification | Letter |
Similar To |