Baxter Healthcare Corporation Recall 87150
Description: Gambro Cartridge Blood Transport System for Hemodialysis Blood Set, product code 109536 - Product Usage: intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems.
Baxter Healthcare Corporation Recall 87150 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1121-2021 |
Event ID | 87150 |
Event Description | Gambro Cartridge Blood Transport System for Hemodialysis Blood Set, product code 109536 - Product Usage: intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems. |
Product Type | Devices |
Distribution | Worldwide distribution. |
Quantity | 211155 devices |
Recall Reason | complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received. |
Device Classification | 20210224 |
Device Code Info | UDI 07332414085216, All lots within expiry |
Center Classification Date | 20210216 |
Recall Initiation Date | 20201230 |
Recalling Firm | Baxter Healthcare Corporation |
Initial Notification | Letter |