SenTec AG Recall 87054
Description: SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous, noninvasive patient monitoring of carbon dioxide partial pressure (PCo2), oxygen partial pressure (POs), functional oxygen saturation (PsO2), and pulse rate (PR), using either a single digital sensor (V-Sign Sensor 2 or OxiVenT sensor).
SenTec AG Recall 87054 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1134-2021 |
Event ID | 87054 |
Event Description | SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous, noninvasive patient monitoring of carbon dioxide partial pressure (PCo2), oxygen partial pressure (POs), functional oxygen saturation (PsO2), and pulse rate (PR), using either a single digital sensor (V-Sign Sensor 2 or OxiVenT sensor). |
Product Type | Devices |
Distribution | IL, PA, FL, NJ, GA, MA |
Quantity | 15 devices |
Recall Reason | Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards. |
Device Classification | 20210224 |
Device Code Info | Serial Numbers: 303170, 311263, 311166, 311408, 303426* *serial number information current as of today's date |
Center Classification Date | 20210218 |
Recall Initiation Date | 20201210 |
Recalling Firm | SenTec AG |
Initial Notification | Letter |
Similar To |