Ortho-Clinical Diagnostics, Inc. Recall 87200
Description: VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.
Ortho-Clinical Diagnostics, Inc. Recall 87200 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1136-2021 |
Event ID | 87200 |
Event Description | VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays. |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada. |
Quantity | 4,062 cartons of 1,000 tips (Totaling 4,062,000 tips) |
Recall Reason | A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays. |
Device Classification | 20210303 |
Device Code Info | Lot 413420 Model(s): VITROS 250, VITROS 350, VITROS ECi, VITROS 5,1 FS, VITROS 3600, VITROS 4600, VITROS 5600, VITROS XT7600. |
Center Classification Date | 20210219 |
Recall Initiation Date | 20210114 |
Recalling Firm | Ortho-Clinical Diagnostics, Inc. |
Initial Notification | Letter |