Product Safety Recalls

Product Recall Tracker

Shanghai United Imaging Healthcare Co., Ltd. Recall 87289

Description: Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Usage: intended to used by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. Not for mammography.

Shanghai United Imaging Healthcare Co., Ltd. Recall 87289 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1160-2021
Event ID87289
Event DescriptionDigitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Usage: intended to used by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. Not for mammography.
Product TypeDevices
DistributionUS Nationwide distribution in the state of TX.
Quantity6 units
Recall ReasonThere is a potential for the X-Ray tube head (Touch screen interface) to perform an uncommanded motion. This could result in a collision with the patient.
Device Classification20210310
Device Code InfoSerial Number: 270012, 270013, 270014, 270015, 270016, 270017
Center Classification Date20210302
Recall Initiation Date20210125
Recalling FirmShanghai United Imaging Healthcare Co., Ltd.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.