Product Safety Recalls

Product Recall Tracker

Siemens Healthcare Diagnostics, Inc. Recall 87319

Description: Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disorders (such as diabetes mellitus), various liver and renal diseases, and in the assessment of risk for atherosclerosis and cardiovascular disease.

Siemens Healthcare Diagnostics, Inc. Recall 87319 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1188-2021
Event ID87319
Event DescriptionDimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disorders (such as diabetes mellitus), various liver and renal diseases, and in the assessment of risk for atherosclerosis and cardiovascular disease.
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the countries of Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Portugal, SA, Slovakia, Slovenia, Spain, Switzerland.
Quantity4174 devices (Domestic: 2836 devices, Foreign: 1338 devices)
Recall ReasonSiemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridges lot 20062BA.
Device Classification20210317
Device Code InfoProduct: DV HDLC; Lot #: 20062BA; UDI Number: 0084276802043820062BA21030210464340840
Center Classification Date20210308
Recall Initiation Date20210203
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Initial Notification Letter
Similar To