Sensus Healthcare, Inc. Recall 91795

Description: Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

Sensus Healthcare, Inc. Recall 91795 Information

StatusTerminated
Mandated?Voluntary: Firm initiated
Recall NumberZ-1235-2023
Event ID91795
Event DescriptionSensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities
Product TypeDevices
DistributionUS Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.
Quantity23 units
Recall ReasonWhen the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.
Device Classification20230315
Device Code InfoSoftware version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272
Center Classification Date20230308
Recall Initiation Date20230210
Recalling FirmSensus Healthcare, Inc.