R & D Systems, Inc. Recall 87223

Description: Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.

R & D Systems, Inc. Recall 87223 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1255-2021
Event ID87223
Event DescriptionQuantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.
Product TypeDevices
DistributionUSA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,
Quantity156 kits
Recall ReasonR&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.
Device Classification20210324
Device Code InfoCatalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604
Center Classification Date20210315
Recall Initiation Date20210120
Recalling FirmR & D Systems, Inc.