R & D Systems, Inc. Recall 87223
Description: Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.
R & D Systems, Inc. Recall 87223 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1255-2021 |
Event ID | 87223 |
Event Description | Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia. |
Product Type | Devices |
Distribution | USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan, |
Quantity | 156 kits |
Recall Reason | R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits. |
Device Classification | 20210324 |
Device Code Info | Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604 |
Center Classification Date | 20210315 |
Recall Initiation Date | 20210120 |
Recalling Firm | R & D Systems, Inc. |
Initial Notification | Letter |
Similar To |