Product Safety Recalls

Product Recall Tracker

Canon Medical System, USA, INC. Recall 87322

Description: Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000F Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

Canon Medical System, USA, INC. Recall 87322 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1271-2021
Event ID87322
Event DescriptionCanon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000F Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.
Product TypeDevices
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.
QuantityTotal of all Systems = 145 systems (U.S.)
Recall ReasonSystem table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Device Classification20210324
Device Code InfoTable CAT-880B Models: INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/BHUP INFX-8000F/AH INFX-8000F/AH INFX-8000F/UU INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/AH INFX-8000F/BH Table Serial Numbers: 99E1582258 99F18Y2428 99C1482200 99E15Y2271 99F1862406 99G19Z2438 99E1662308 99E1762365 99A1337005 99F1852401 99E1612286 99E1712347 99E1712345 99C14Z2224 99E1592262 99G1972436 System Serial Numbers: AHA1582003 AHC18Y2013 AHA1482001 AHA15Y2005 AHC1862012 BHB19Z2002 AHB1662007 AHB1762010 UUB12X2034 AHC1852011 AHA1612006 AHB1712009 AHB1712008 AHA14Z2002 AHA1592004 BHB1972001
Center Classification Date20210317
Recall Initiation Date20201208
Recalling FirmCanon Medical System, USA, INC.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy