Product Safety Recalls

Product Recall Tracker

ConvaTec, Inc Recall 93896

Description: ConvaTec Natura Stomahesive Post-Operative / Surgical Ostomy System, REF 416938, 1 3/4 in. 45mm, Sterile.

ConvaTec, Inc Recall 93896 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1276-2024
Event ID93896
Event DescriptionConvaTec Natura Stomahesive Post-Operative / Surgical Ostomy System, REF 416938, 1 3/4 in. 45mm, Sterile.
Product TypeDevices
DistributionDistribution was made to AK, CA, FL, IL, IN, MI, NE, NJ, NY, OH, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Canada.
Quantity1,715 units
Recall ReasonThe kit contains a flange that has a 45mm coupling ring, however, the pouch within the kit is for a 57mm coupling ring.
Device Classification20240313
Device Code InfoLot #2C02496, UDI-DI #00768455116096.
Center Classification Date20240307
Recall Initiation Date20240117
Recalling FirmConvaTec, Inc
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.