RAYSEARCH LABORATORIES AB Recall 91867
Description: Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs
RAYSEARCH LABORATORIES AB Recall 91867 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1281-2023 |
Event ID | 91867 |
Event Description | Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs |
Product Type | Devices |
Distribution | US Nationwide |
Quantity | 10770 systems |
Recall Reason | It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle. |
Device Classification | 20230329 |
Device Code Info | UDI-DI: 0735000201067920221007; |
Center Classification Date | 20230322 |
Recall Initiation Date | 20230217 |
Recalling Firm | RAYSEARCH LABORATORIES AB |
Initial Notification | Letter |