Mizuho OSI Recall 87318
Description: Mizuho CE IPX4, mode MOT-5602BW, 100-240, 50-60Hz, 400VA, REF 18-140-20 containing hand control Unit - Product Usage: intended to support a patient during surgical operations.
Mizuho OSI Recall 87318 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1293-2021 |
Event ID | 87318 |
Event Description | Mizuho CE IPX4, mode MOT-5602BW, 100-240, 50-60Hz, 400VA, REF 18-140-20 containing hand control Unit - Product Usage: intended to support a patient during surgical operations. |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the state of CA, and the countries of Costa Rica, Chile, Bolivia, Peru. |
Quantity | 26 units |
Recall Reason | There is a potential that hand control units may cause the tabletop to slide in the opposite direction from that selected by the user. This could result in a delay in procedure. |
Device Classification | 20210331 |
Device Code Info | S/N (Control unit) - S/N (Operating table): 10290134 - MEC-20195-0034, 10290135 - MEC-20195-0036, 10290136 - MEC-20195-0035, 10290137 - MEC-20195-0033, 10490143 - MEC-20198-0038, 10490144 - MEC-20198-0037, 10490145 - MEC-20198-0040, 10490147 - MEC-20198-0039, 10690148 - MEC-2019Y-0054, 10690149 - MEC-20199-0049, 10690150 - MEC-2019Y-0059, 10690153 - MEC-2019Y-0058, 10690154 - MEC-20199-0046, 10690155 - MEC-2019Y-0057, 10690156 - MEC-20199-0050, 10690157 - MEC-20199-0044, 10690159 - MEC-20199-0047, 10690160 - MEC-20199-0051, 10690161 - MEC-20199-0045, 10690162 - MEC-20199-0048, 10690163 - MEC-2019Y-0053, 10790167 - MEC-2019Y-0052, 10790169 - MEC-2019Y-0056, 10790170 - MEC-2019Y-0055, 211080129 - MEC-2018Z-0029, 11080130 - MEC-2018Z-0028 |
Center Classification Date | 20210323 |
Recall Initiation Date | 20210208 |
Recalling Firm | Mizuho OSI |