Baxter Healthcare Corporation Recall 90322

Description: NaviCare Nurse Call/Voalte Nurse Call

Baxter Healthcare Corporation Recall 90322 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1306-2022
Event ID90322
Event DescriptionNaviCare Nurse Call/Voalte Nurse Call
Product TypeDevices
Distributionnationwide and Canada
Quantity283 installations
Recall ReasonAn issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Device Classification20220706
Device Code Infosoftware versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300
Center Classification Date20220630
Recall Initiation Date20220531
Recalling FirmBaxter Healthcare Corporation