Mckesson Medical-Surgical Inc. Corporate Office Recall 94024

Description: Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"

Mckesson Medical-Surgical Inc. Corporate Office Recall 94024 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1313-2024
Event ID94024
Event DescriptionSteris Corporation Endo Cleaning Indicator VERIFY" RESITEST"
Product TypeDevices
DistributionUS Nationwide distribution in the states of AZ, PA and TX.
Quantity25 units
Recall ReasonFacility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
Device Classification20240327
Device Code InfoCatalogue Number: LCC101; UDI/DI: 10724995220072
Center Classification Date20240319
Recall Initiation Date20240109
Recalling FirmMckesson Medical-Surgical Inc. Corporate Office