Philips North America, LLC Recall 87418
Description: Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481
Philips North America, LLC Recall 87418 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1326-2021 |
Event ID | 87418 |
Event Description | Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481 |
Product Type | Devices |
Distribution | International distribution in the country of China. |
Quantity | 31773 units |
Recall Reason | Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit |
Device Classification | 20210407 |
Device Code Info | Software version: SW A.02.63 |
Center Classification Date | 20210330 |
Recall Initiation Date | 20210323 |
Recalling Firm | Philips North America, LLC |
Initial Notification | Letter |
Similar To |